Download HAMEG HM OSCILLOSCOPE UM service manual & repair info for electronics experts. Find great deals for HAMEG HM Oszilloskop 20mhz digital Storage. Shop with confidence on eBay!. I have been experiencing a problem with my Hameg HM oscilloscope. It was working correctly then the trace suddenly went back to the left.

Author: Tygojinn Fenririsar
Country: Lesotho
Language: English (Spanish)
Genre: Relationship
Published (Last): 27 June 2004
Pages: 254
PDF File Size: 16.93 Mb
ePub File Size: 17.61 Mb
ISBN: 258-4-45211-268-9
Downloads: 67487
Price: Free* [*Free Regsitration Required]
Uploader: Mezshura

If the signal frequency is not known, there are two other methods for correct time coefficient setting.

Thus any intermediate value inside the 1 scale can be set. The timebase fine control knob is then out of operation. Correction of a few degrees is pos- sible by turning a potentiometer behind an opening on the front panel marked with TR using a small screw driver.

HAMEG Hm208 Oscilloscope 20mhz Digital Storage

Battery-powered hk AC supply operated optional. Do not depress any further buttons. The condition for this is that no modifications have been made in the instrument. Using the 2 V CAL. The time resolution poses no problems. This must be set either to DC or to LF. The clear advantage over ordinary probes are field replaceable parts and the HF-compensation feature on the 1 0: Switch test signal coupling to AC or DC.

When using attenuators probes, even high internal imped- ance sources are only slightly loaded approx. I button may be used. For additional information, see the Global Shipping Program terms and conditions – opens in a new window or tab This amount includes applicable customs duties, taxes, brokerage and other fees.

  ETRALI MACH 3D PDF

With the time axis expanded fold X-MAG. Amplitude measurement with Y fine control at right stop CAL. If they are depressed, the memory contents are retained until the mains voltage is switched off or fails. All operating modes, such as single- or dual-channel operation, add- or dif- ferential mode, X- and Y-expansion, X-Y operation, as well as the various trig- ger facilities are also available in storage operation.

When the instrument is assembled later, pay attention that the casing is seated behind the edge of the front molding lip on all hame. All four buttons out for mono mode. New storage is only possible on each channel when the relevant button is released out. Neither overshooting nor rounding off is allowable. Without a test signal applied, the base line is only vis- ible if this button is not pressed in the AT-position automa- tic triggering.

Oszillograph HM208

The leading edge should be as steep as possible but the top as straight as possible. The 1 MQ input resistance can equally serve as R here, so that only a suitable capacitor C needs to be connected in series. The current time sweep is inter- rupted. Adjusts vertical position of trace for Channel 1 and Channel II. The faster line pulses are then attenuated by a low pass filter to such an extent it is easily possible to trigger on the leading or trail- ing edge of the frame-synchronizing pulse using the level control.

  LAS VIEJAS TRAVESTIS COPI PDF

For amplitude measurements, the variable control on the input attenuator switch yameg be in its calibrated position CAL.

The variable controls for amplitude and time base and the LEVEL knob can also be used for this adjustment – without influence on the result.

The test methods described in the test plan of this manual can be performed without great expenditure on measuring instru- ments. With AC coupling of low frequency signals, the attenuation is no longer independent of frequency, pulses can show pulse tilts.

Hameg HM 2 Kanal Storage | eBay

Ml 2 Subject to change without notice Automatic triggering can be used both with internal and external via the socket TRIG. To obtain a well-triggered display in hameeg case, the assistance of the variable time control may be required.

DC-input coupling is therefore of quite special importance with a xlOO attenua- tion probe which usually has a voltage rating of max.

See all condition definitions – opens in a new window or tab Phase difference measurement in DUAL mode A larger phase difference between two input signals of the same frequency and shape can be measured very sim- ply on the screen in Dual mode DUAL button depressed.